ALF245 Axial compensated load cell

Loadcell ALF245

The ALF245 is ideally suited to low range engineering force measurements and process weighing. When precision and easy installation are required various configurations allow the load cell to be used in both tensile and compressive applications.

Geometry:
Flexure strain assembly in cylindrical housing, open or weather sealed with end internal fixing. For universal use in tension and compression, with compensation for off axis load inputs.

Technical data

Parameter Value

Rated load:

100 N / 200 N / 400 N / 800 N

Non-linearity, terminal:

±0.1 % RL

Hysteresis:

±0.1 % RL

Creep, 20 min:

±0.1 % AL

Repeatability:

±0.02 % RL

Rated output, nominal:

2.2 mV/V

Rated output, rationalised:

2.0 mV/V ±0.1 % RL

Zero load output:

±4 % RL

Temperature effect on rated output:

±0.002 % AL/K

Temperature effect on zero load output:

±0.005 % RL/K

Compensated temperature range:

-10 ... +50 °C

Operating temperature range:

-10 ... +80 °C

Supply voltage, recommended:

10 V

Supply voltage, max.:

10 V

Bridge resistance:

350 Ω

Insulation resistance, minimum at 50 VDC:

500 MΩ

Overload, safe:

120 % RL

Overload, ultimate:

200 % RL

Weight (excl. cable):

approx. 40 g to 50 g

Material:

ranges 100 N and 200 N: aluminium body and stainless steel case
ranges 400 N and 800 N: stainless steel

Rated load Structural stiffness, nom. Rated load Structural stiffness, nom.

100 N

3.2 x 106 N/m

400 N

3.0 x 106 N/m

200 N

6.4 x 106 N/m

800 N

6.0 x 106 N/m

Notes:

  1. RL = rated load
  2. AL = applied load
  3. Temperature coefficients apply over the compensated range

Dimensions

Loadcell  ALF245 dimensions

Dimensions in ''mm'', approx. values. These drawings are for information only and not intended for construction purpose.

Electrical Connections

red

+ supply voltage

blue

- supply voltage

yellow

+ output signal

green

- output signal

orange

screen

The load cell is fitted with 2 m of PVC insulated 4 core screened cable type 7-2-4C. Reverse the signal connections to obtain a positive signal in tension mode. The screen is not connected to the load cell body.

Ordering Information

unrationalised: rationalised:

ALF245CF00H0

compression

 

ALF245CF00HN

compression

ALF245TF00H0

tension

 

ALF245TF00HN

tension

ALF245UF00H0

bi-directional

 

ALF245UF00HN

bi-directional

Change the first 0 to an R for the IP65 version.
Please add range in the required units.

Safety note:
When using the load cell in tension mode it is essential to provide additional safety precautions like safety chains etc. for catching the load in a breakage, which cannot be excluded completely.