AS30 | AS50 Precision Thin-film Pressure Sensors

Pressure Sensors AS30 and AS50

For applications that need consistent high levels of performance, reliability and stability the AS30/AS50 series sputtered thin film units offer unbeatable price performance ratio in a small package size with all stainless steel wetted parts in the volumes required. A wide choice of electrical outputs as well as both electrical and pressure connections means the unit is suitable for most applications without modification. The compact construction makes it ideal for installation where space is at a premium.

The sensors are available in series AS30 with pressure signal and in series AS50 with pressure and temperature signal (dual output).

Features

  • Superior long-term stability
  • High accuracy
  • Rugged design
  • Wetted part in stainless steel
  • Low cost

Applications

  • Hydraulic controls
  • HVAC systems
  • Engine controls (incl. Diesel)
  • Industrial process controls
  • Off-highway equipment controls
  • Compressors and pumps

Technical dataAS30 | AS50

Pressure range in [bar]

0 … 4

5

10

16

25

40

60

100

160

250

400

600

1000

1600

 

vented gauge

sealed gauge

Proof pressure [factor FS]

3x

3x

3x

3x

3x

2x

2x

2x

2x

2x

2x

2x

2x

1.4x

Burst Pressure [factor FS]

40x

40x

40x

40x

40x

20x

20x

20x

10x

10x

10x

4x

4x

1.8x

Fatigue life

Designed for more than 100 million cycles

Long-term drift

0.2 % FS/year (non-cumulative)

Accuracy of pressure output

0.25 % FS

Accuracy of temperature output (AS50 only)

3.5 % of temperature span

Thermal error, typical

1.5 % FS/100 °C

Bandwidth

AS30: DC ... 250 Hz
AS50: DC ... 1 kHz

Compensated temperature range:

-40 ... +105 °C

Operating temperature range

-40 ... +105 °C

Zero tolerance

0.5 % of span

Span tolerance

0.5 % of span

Pressure port

see table below

Wetted parts material

17-4 PH stainless steel

Electrical connection

see table below

Enclosure

IP65 for electrical connection codes: 1
IP67 for electrical connection codes: 2, 3, 4, 5, 6, 7

Vibration

40 g peak-peak, (vibration 20 … 1000 Hz sinusoidal with 40 g peak per MIL-STD-810E)

Shock

Withstands free fall to EC 68-2-32 procedure 1

EMV

100 V/m (radiated immunity)

Approvals

CE, fully RoHS compliant

Weight

35 grams

Individual Specifications

Voltage output:

Output signal AS30

1 … 5 V, 1 … 6 V, 0 … 5 V, 0 … 10 V, 0.5 … 4.5 V (current 5.5 mA max.)

Output signal AS50

1 … 5 V, 1 … 6 V, 0.5 … 4.5 V (current 5.5 mA max.)

Supply voltage

2 V above full scale to 30 VDC max. at 5.5 mA max.

Source and sinks

2 mA max.

 

Current output (4 … 20 mA, 2-wire):

Output signal (AS30 only)

4 … 20 mA

Supply voltage

8 … 30 VDC (24 VDC max. for temperature 100 °C and above)

Max. load resistance

(Supply voltage - 8) x 50 Ohm

Min. load resistance

(Supply voltage - 24) x 50 Ohm

 

Ratiometric output:

Output signal

0.5 … 4.5 V at 5.5 mA

Supply voltage

5 VDC ±10 %

Pressure Ports

Code 01

Code 02

Code 04

Code 05

 

 

G1/4’’ male

G1/4’’ A
integral face seal

7/16’’-20 UNF with 37° flare

¼’’-18 NPT

 

 

Code 06

Code 07

Code 08

Code 09

Code 10

M12 x 1.5

M12 x 1.5 HP
metal washer seal

7/16’’-20 UNF
O-ring

1/8’’-27 NPT

7/16’’ Schraeder

Dimensions in "mm", all values are approx. values
These drawings are for information only and not intended for construction purpose.
Please contact us for detailed drawings.

Electrical Connections

DIN43650C (Code 1) M12 x 1P (Code 2) Deutsch DT04-4P (Code 5)
Pin Voltage Current
1 + signal pressure N/A
2 + supply voltage +
3 + signal temp. N/A
4 ground -
Pin Voltage Current
1 + supply voltage +
2 + signal pressure N/A
3 ground -
4 + signal temp. N/A
Pin Voltage Current
1 ground -
2 + supply voltage. +
3 + signal temp. N/A
4 + signal pressure N/A
Amp Superseal 1.5 (Code 3) Packard MetriPack (Code 6) Integral Cable (Code 7)
Pin Voltage Current
1 + signal pressure N/A
2 ground -
3 + supply voltage N/A
Pin Voltage Current
A ground -
B + supply voltage. +
C + signal pressure N/A
Colour Voltage Current
red + supply voltage +
black ground -
white + signal pressure N/A
green + signal temp. N/A

Hex is 22 mm
More ports on request.
Dimensions in „mm“, all values are approx. values
These drawings are for information only and not intended for construction purpose.
Please contact us for detailed drawings

Ordering Information

AS3

x

yy

z

0

MB

KA

AS5

x

yy

z

t

MB

KA

 

Output Supply voltage x=

1 ... 5 V

7 ... 30 VDC

2

0.5 ... 4.5 V

6.5 ... 30 VDC

3

0.5 ... 4.5 V ratiometr.

5 VDC

4

1 ... 6 V

8 ... 30 VDC

5

0 ... 5 V

7 ... 30 VDC

6

0 ... 10 V

12 ... 30 VDC

7

4 ... 20 mA 2-wire

8 ... 30 VDC

8

 

Pressure Ports: yy=

G1/4 external, hex22, for seal ring

01

G1/4 external with integral O-ring, hex22

02

M14 x 1.5 per ISO 6149-2

03

7/16’’-20 UNF, 37° flare (SAEJ514), hex22

04

1/4’’-18 NPT, hex22

05

M12 x 1.5 (6G), hex22, for O-ring

06

M12 x 1.5 HP MB ?1000 bar, hex22

07

7/16’’-20 UNF for O-ring

08

1/8’’-27 NPT, hex22

09

7/16’’ Schraeder

10

M14 x 1.5 external with 60° cone

11

 

Electrical Connections z=

Mini-DIN43650C (4-pin)

1

M12 x 1 (4-pin)

2

Amp Superseal (3-pin)

3

Deutsch DTD4-4P (4-pin)

5

Packard Metri-Pack150 (3-pin)

6

integral cable

7

 

Temperature Output (1) ( AS50) t=

-40 °C ... +105 °C

T1

0 °C ... +100 °C

T2

0 °C ... +80 °C

T3

 

Ranges MB=

4 bar V

BV0004 (3)

5 bar V

BV0005

10 bar V

BV0010

16 bar V

BV0016

25 bar V

BV0025

40 bar V

BV0040

60 bar V

BV0060

100 bar V

BV0100

160 bar S

BS0160

250 bar S

BS0250

400 bar S

BS0400

600 bar S

BS0600

1000 bar S

BS1000 (2)

1600 bar S

BS1600 (2)

V = vented to atmosphere
S = sealed

Cable Length KA =

no cable

00

1 m

01

2 m

02

3 m

03

4 m

04

5 m

05

Notes:

  1. Model AS50 available with voltage outputs 2, 3, 4 and 5 and limited to connections that have 4 pins like electrical codes 1, 2, 4 and 5. Accuracy is 3.5 % of temperature span. Requires additional 2 mA of power.
  2. Ranges 1000 bar and above available with pressure port 07 (M12 x 1.5 HP) only
  3. Range 4 bar not available with 4 .. 20 mA and 0 .. 10 V outputs.



To create a code for a customized version, we will use the code for the nearest standard version and add a -Zxxx number that defines exactly the specs of the customized version!

For sensors with individual calibration certificate, a "K" has to be added to the sensor code. (additional charge!)

Please order all mating connectors and cables separately!

The specifications are subject to change without prior notice.