Omniprobes

For measurement applications involving higher angularity, unknown angularity, or reversed flows


About the Omniprobes

These probes were designed and patented for applications requiring a very high range of incidence angles. As their name suggests, they are capable of near-omni-directional flow measurement. Omniprobes are also ideal in applications where the predominant direction of the flow is not known in advance or where the probe may see a complete reversal of the flow during testing. A micro-machined spherical tip accommodates 12 pressure ports that allow for measurement of +/- 150 degrees of flow angle.

 

Key features Omniprobes

  • 12 hole
  • Spherical tip geometry
  • Standard tip diameter 6.35 and 9.53 mm
  • Capable of measuring flow vectors up to ±150 °
  • Temperature ratings up to 500° C
  • High accuracy NIST traceable aerodynamic calibrations
  • Up to 9000 discrete aerodynamic calibration points per speed
  • 303, 304 Stainless steel construction; Optional nickel based super alloy for high strength
  • Custom engineering solutions available
  • Available in Straight and L-shaped geometries
  • Standard media: non-reactive gases, air, water, and other fluids
  • Angle error less than ±2°
  • Velocity error less than ±3% or ±0.005 m/s (whichever is larger)

 

Applications Omniprobes

  • High-angularity wake measurements around automobile wheels
  • Wake measurements behind automobiles and rolling-road wind tunnels
  • Wind engineering or combustor swirl flows

Avez-vous besoin de ce capteur pour votre prochain projet?

Demandez un devis dès aujourd'hui ou laissez vos coordonnées pour recevoir plus d'informations de la part de l'un de nos experts.

Parler à un expert

Faites-le à votre façon

Vous ne trouvez pas exactement ce que vous cherchez ? Nous proposons des capteurs et des systèmes de mesure standard, mais nous pouvons également vous aider à concevoir un capteur sur mesure ou une solution de mesure complète. N'hésitez pas à nous contacter pour obtenir des conseils ou un devis.

Parler à un expert

Spécifications techniques

Spécifications Valeur
Average measured angular deviation <1°
Average measured velocity deviation ± 3% or ±1 m/s (whichever is larger)
Probe calibrations 5 m/s to Mach 1.0
Temperature ratings to 900°C
Tip diameters 6.35 mm