HL-Z Wet Differential Pressure Transducers /High Line Wet

The High Line Wet/Wet Differential Pressure Sensor is designed to accept extreme line pressures of up to 5000 psi.

About the HL-Z Wet Differential Pressure Transducers /High Line Wet

The High Line Wet/Wet Differential Pressure Model HL-Z is designed to accept extreme line pressures of up to 5000 psi.* This unique design features standard bonded foil strain gages on stainless steel. Liquid and gas differential pressures of 50 psid to 7500 psid are measured within a 0.25 % to 0.50 % full scale accuracy. Typical applications include flow measurement and laboratory testing.

 

Key features HL-Z

  • Line pressure to 5000 psi
  • 0.25 % accuracy
  • Intrinsically safe available (2N option only)
  • CE approved

 

Applications HL-Z

  • Flow measurement
  • Laboratory testing

Specifications

Pressure ranges
50, 75, 100, 150, 200, 300, 500, 750, 1000, 1500, 2000, 3000, 5000, 7500 psid (±3.5 to ±500 bar)
Accuracy
±0.25 % full scale
Linearity
±0.15 % full scale (typical)
Hysteresis
0.10 % full scale (typical)
Non-repeatability
±0.05 % full scale (typical)
Output (standard)
2 mV/V (nominal)
Line pressure
2500 psi
Resolution
infinite
Temperature, operating
-54 °C to 121 °C [-65 °F to 250 °F]
Temperature, compensated
15 °C to 71 °C [60 °F to 160 °F]
Temperature, effect, zero
±0.5 % full scale/100 °F
Temperature, effect, span
±0.5 % reading/100 °F