CTL S-beam Loadcell

Protection class IP 68

Features CTL

  • Range: 100, 200, 300, 500, 100, 2500, 5000, 7500, 10000, 12500kg
  • 17-4 PH stainless steel construction
  • Combined error < +/- 0.02 %
  • Protection rating IP 68 (IPX9K)
  • Accuracy class C3 (OIML R60)

Specifications

Measurement range
100, 200, 300, 500, 100, 2500, 5000, 7500, 10000, 12500kg
Rated output
2 mV/V +/- 0.1 %
Temperature effect on zero
0.005 % / °C
Temperature effect on span
0.003 % / °C
Compensated temperature range
- 10°C / + 50°C
Operating temperature range
- 20°C / + 70°C
Creep at nonimal load after 4 hours
0.05 %
Max supply voltage without damage
15 Volt
Input resistance
350 ohm +/- 5
Output resistance
350 ohm +/- 2
Zero balance
+/- 1 %
Insulation resistance
> 5000 Mohm
Safe overload (% of Full Scale)
150 %
Ultimate overload (% of Full Scale)
> 300 %
Deflection at nominal load
0.3 mm