SG-2K-IP Strain Gauge Amplifiers
Dual channel strain gauge amplifier
About the SG-2K-IP Strain Gauge Amplifiers
The dual-channel strain gauge measuring amplifier SG-2K-IP allows supply and signal evaluation of two connected transducers with a strain gauge full bridge greater than 300 ohms. Input and output are galvanically isolated. The connection is done in 4-wire technology. For further evaluation standard analogue outputs are available. The measuring amplifier is built-in into a robust EMC die-cast aluminium enclosure (IP65) which is suited for application in rough and industrial environments.
The amplification/channel of the amplifier can be adapted by an internal precision resistance.
The potentiometers Zx (Zero) and Gx (Gain) which are accessible after removing the enclosure cover allow a calibration correction for each channel.
By the means of an internal dip switch/channel a change of the range of the potentiometer Zx (zero) can be reached. To allow a zero range shift a basic load / tare can be suppressed by a resistance electrically.
Key features SG-2K-IP
- Supply voltage 10 - 18 VDC / 18 - 30 VDC
- Analogue output 0 - 10 V / ±10 V / 4 - 20 mA
- EMC Aluminium die-cast enclosure (IP65)
- Dimensions (W x H x D) 220 x 140 x 70 mm
- Galvanically isolated
Avez-vous besoin de ce capteur pour votre prochain projet?
Demandez un devis dès aujourd'hui ou laissez vos coordonnées pour recevoir plus d'informations de la part de l'un de nos experts.
Téléchargements de produits
Faites-le à votre façon
Vous ne trouvez pas exactement ce que vous cherchez ? Nous proposons des capteurs et des systèmes de mesure standard, mais nous pouvons également vous aider à concevoir un capteur sur mesure ou une solution de mesure complète. N'hésitez pas à nous contacter pour obtenir des conseils ou un devis.
Parler à un expertSpécifications techniques
Spécifications | Valeur |
---|---|
Analogue output | 0 - 10 V, ±10 V, max. 1 mA (short-period short-cicuit proof) 4 - 20 mA, max. 500 Ω |
Dimension (b x h x d) | 220 x 140 x 70 mm |
Electrical connection | EMC cable gland on internal terminal block |
Enclosure | EMC aluminium die-cast enclosure (IP65) |
Input resistance | >3 MΩ |
Isolating proof voltage input to output | 200 V (higher isolated proof voltage on request) |
Limit frequency (-3 db) | 1 kHz, optional up to 30 kHz |
Max. input sensitivity | 100 mV/V at 10 VDC excitation supply |
Non-linearity | ±0.05 % FSO |
Number of measuring channels | 2 (full bridge resistance >300 Ω) |
Power consumption | max. 5 W |
Strain gauge excitation supply | +5 VDC, +10 VDC |
Supply voltage | 10 - 18 VDC, 18 - 30 VDC, electronics protected against reversal voltage |
Temperature, operating | -20 to +50 °C |
Temperature, storage | -20 to +60 °C |
Weight | approx. 1500 g |