APS Absolute Pressure Transducer

Low-range pressure sensor

About the APS Absolute Pressure Transducer

The pressure transducers of the DPS type series are suitable for detection of positive and negative gauge pressure or differential pressure of non-aggressive gases. Optional there is an absolute pressure version APS available. These robust versions can be applied in laboratories and under industrial conditions. Important criteria such as long-term stability, linearity and good reproducibility are guaranteed through their sturdy mechanical construction. Temperature drift is minimised through targeted compensation of every sensor. The wear-free inductive measurement system allows for almost maintenance-free operation.

The integrated electronic system provides a pressure-proportional voltage signal output of 0 … 10 V (option: current signal 0(4) …20 mA). This allows an interference-free transmission of signals even over comparatively large distances. For strongly varying pressures a dampening function is included.

 

Key features APS

  • Measurement ranges from 0.1 mbar to 1000 mbar
  • Series DPS: positive and negative gauge pressure, differential pressure, optional bi-directional
  • Series APS: absolute pressure

Applications APS

  • HVAC
  • Clean room technology
  • Medical technology
  • Filter technology
  • Finishing pass measurement
  • Level measurement (bubbling-through method)
  • Flow velocity measurement (pitot tube, orifice disk)

Specifications

Measurement Range
barometric: 900 - 1100 mbar, 800 - 1200 mbar; absolute: 0 - 1000 mbar, 0 - 500 mbar
Accuracy
1.0%
Output
0 - 10 V [19 - 31 VDC], ±5 V [19 - 31 VDC], ±10 V [19 - 31 VDC], 0 - 20 mA [19 - 31 VDC], 4 - 20 mA [19 - 31 VDC], 4 - 20 mA 2w [12 - 32 VDC]
Priciple
Inductive
Temperature
Specified range +10 to +50 °C
Storage temperature
-10 to +70 °C
Humidity
Up to 80 % relative humidity
Long-term stability
±0.5 % per year (typical)
Housing
Material ABS
Pressure connections
6.6 diameter x 11 mm (for flexible hoses 6 diameter)
Electrical connections
Cable gland M12 x 1.5, screw terminals 1.5 mm² max.
Protection class
II
Sealing
IP54
Standards
EN 61010-1; 2. (08/2002), EN 61000-6-3 (05/2006), EN 61000-6-1 (10/2007), EN 61000-6-2 (03/2006)
Weight
Approx. 0.3 kg (approx. 0.4 kg incl. power pack)
Shock resistance
10 g
Sensor volume
Approx. 3 ml
Volume increase
Approx. 0.2 ml at rated pressure