aCTA Hot-Wire Bridge

Hot-wire bridge for conducting dynamic flow velocity measurements in air.


Key features

  • Dynamic measurements of flow velocities
  • Velocity range: 0.1 – 50 m/s in air, higher velocities and other gases possible
  • Measurement accuracy: 0.1 m/s + 0.5% FS (v < 20 m/s) / 1% FS (20m/s ≤ v < 50 m/s)
  • Different hot-wire probes are available
  • Plug & play
  • Acoustic analysis via headphone

About the aCTA Hot-Wire Bridge

The aCTA hot-wire bridge is a powerful measuring device for precise and dynamic acquisition of air flow velocities. It has been specifically developed for demanding applications in flow measurement technology and stands out due to its high accuracy, ease of use, and versatile application possibilities.

With a calibrated measurement range of up to 50 m/s, the aCTA is particularly suitable for analyzing low to medium flow velocities. Higher velocities can also be measured with appropriate calibration. The system operates based on the proven hot-wire principle, enabling highly sensitive and fast detection of even the smallest flow fluctuations.

A key advantage of the aCTA is its plug-and-play functionality: the device is already optimally matched to hot-wire sensors of type HWP10/xx, eliminating the need for complex adjustments and setup. This saves time and allows immediate use in laboratories or test benches.

The output signal is provided as an analog voltage (±10 V) and can be directly converted into flow velocities, taking calibration and temperature into account. An integrated low-pass filter (5 kHz) reduces unwanted aliasing effects during digital data acquisition, while an unfiltered output allows measurements up to 20 kHz.

An innovative feature is the acoustic signal output: flow fluctuations can be made audible via headphones. This intuitive function significantly facilitates analysis and troubleshooting, especially for acoustically noticeable flow phenomena.

Power is supplied flexibly via a plug-in power supply or an external DC source in the range of 9–18 V, making the aCTA suitable for both stationary and mobile applications.

Key benefits at a glance:

  • Precise measurement of flow velocities up to 50 m/s
  • Plug-and-play operation without calibration effort
  • Analog output (±10 V) for easy processing
  • Integrated low-pass filter and high-speed measurement option
  • Acoustic analysis of flow signals
  • Flexible power supply (9–18 V)

The aCTA hot-wire bridge is the ideal solution for engineers, researchers, and developers who require reliable and high-resolution flow measurements—from basic research to industrial applications.

Do you need this sensor for your next project?

Request a no-obligation quote or send us your contact details for a free, no-obligation consultation. Our sales engineers will then contact you within a maximum of 1 to 2 working days.

Talk to an expert

Product downloads

Talk to our experts

Not exactly what you are looking for? We offer standard sensors, but can also help you with sensors adapted for you or a fully customised measurement solution. Contact us for advice or pricing.

Talk to an expert

Technical Specifications

Specification Value
Velocity range 0.1 - 50 m/s, higher velocities possible
Frequency resolution 0…5 kHz with 3 dB lowpass, unfiltered up to about 10 kHz
Fluid Air, other non-corrosive gases are possible but must be calibrated separately no particles, no air pollution
Temperature 0… 50°C within factory calibration, -10…70°C possible
Humidity Non-condensing
Velocity 0…20 m/s 0.1 m/s + 0.5% of reading
Velocity 20…50 m/s 1% full-scale
Supply 9…18 V DC, min. 0.5 A, 10 W
Signal ±10 V
Headphone min. 32 Ohm
Supply / signal / enable Binder 711/712, 4 pin
Signal BNC
Ambient conditions for electronics Temperature 5° C…40°C / Humidity Non-condensing

Related products

Need something else?

eCCT Hot-Wire Bridge

  • Dimensions: 112 mm x 27 mm x 17 mm
  • Humidity: Non-condensing
  • Temperature: 5° C…40° C
Datasheet View product
4-channel data acquisition module

4CTA Hot-Wire Measurement System

  • Data transfer via USB
  • Software included
  • Free analogue and digital in- and outputs onto customer specific connectors
Datasheet View product
Data acquisition device with multiple channels

sDAQ-Set Data Acquisition System

  • 1x PT100 temperature sensor
  • 2x ePressure differential pressure sensor
  • Integrated barometrical pressure sensor
Datasheet View product